To insert this chart into a binder punch out the holes and fold the chart as illustrated To use this chart as a wallchart cut off the top of the chart as illustrated BRONCHOSCOPE DISINFECTION GUIDE EXERA BF 160 Series and EVIS BF 240 Series Bronchovideoscopes Refer to the companion BRONCHOSCOPE CLEANING GUIDE for detailed cleaning information WARNING This guide is only a summary of the steps necessary to properly disinfect your bronchoscope Be sure to follow the detailed steps outlined in the ENDOSCOPE REPROCESSING MANUAL that was included with your bronchoscope when purchased HIGH LEVEL DISINFECTION Suction Cleaning Adapter Fill basin with disinfectant Attach a suction cleaning adapter MAJ 222 to the previously cleaned broncho sco ...