Taylor - TE10SSW
Below are all the different types of Taylor. Select the device of your choice, to download the manual
Click here to download the manualContents
Calibration Function Occasionally the scale s accuracy should be verified by measuring a known weight of the scale s approximate capacity Tare the scale by pressing ZERO SET place the weight on the platform and note the reading Calibrate the scale if necessary Calibration Procedure You must use a 5 lb or a 2000g calibrated weight ONLY to properly recalibrate the scale 1 Remove any weight from the scale platform 2 Turn the scale on Select the weight unit to be calibrated lb or g 3 Turn the scale off The weight unit selected will be stored in memory 4 Press and hold down the TARE MODE button While holding down TARE MODE also press the ON OFF button The displays shows CAL then zero Release both buttons 5 Press the T ...